Specifications
Reading |
Ocular (AR4) |
|
Digital display (AR4D) |
Measuring range |
1,3000-1,7000 nD |
|
0-95 %Brix |
Accuracy |
0,0002 nD |
|
0,1 %Brix |
Scale division |
0,0005 nD |
|
0,25 %Brix |
Power supply |
110 V or 220 V |
|
(switchable) |
Temperature measured |
Digital thermometer -50 to 200 °C |
Illumination |
Scale illumination and |
|
LED-illumination (590 nm) for measuring prism |
Dimensions |
140 X 100 X 235 mm (AR4) |
|
230 x 110 x 270 mm (AR4D) |
Weight |
4.4 kg (AR4) |
|
5.5 kg (AR4D) |
Specials |
Adjustable scale |
|
Temperature controlled prism |
|
Complies with ASTM D1218 |
|
Complies with ISO 489 |
Applications
Mixing-ratio control, quality and quantity control in following industries:
- Pulp and paper industry
- Chemical industry
- Beverage industry
- Food industry
- Sugar and sweetener industry
- Textile industry
Scope of Supply
- Digital thermometer -50 to 200 °C
- Calibration standard with defined index of refraction
- Contact liquid
- Adjusting tool
- Dust protection cover
Accessories
Illumination |
AR31 |
for AR3 and AR4 |
|
AR32 |
for AR4D |
Calibration |
RI34 |
Calibration liquid 1.3400 nD |
|
RI39 |
Calibration liquid 1.3900 nD |
|
RI43 |
Calibration liquid 1.4300 nD |
|
RI48 |
Calibration liquid 1.4800 nD |
|
RI65 |
Calibration liquid 1.6500 nD |
|
RK01 |
Calibration standard |
Prisms |
AR11-3 |
Measuring prism |
|
AR12-3 |
Illumination prism |
Miscellaneous |
AR18 |
Digital thermometer |
Thermostats |
ARFT |
Filter for thermostats |
|
DC10-K10 |
Thermostat with Cooling |
|
DC10-P5-U |
Thermostat without Cooling |
|
PT31 |
Peltier thermostat |
|
PT35 |
Thermostat cleaner |
|
|
|
Abbe refractometers are easy to use and only a small sample volume is required. They are well suited for all kinds of samples, no matter if solid, viscous or fluid. Transparency does not affect measurement. Refractive index (nD), dispersion (nf-nC) and concentration (%Brix) can be measured. For better accuracy prisms can be tempered and a digital thermometer is included in delivery.