Digital refractometers DR6000-TF fully automatic

Price on request

6.978,16 10.700,48 

incl. VAT

+ Shipping costs

Proven Peltier measurement technology: the fully automatic measurement process increases productivity and cost-effectiveness

Item number N/A Category

In addition to the Peltier element, the DR6000-TF models have a flow-through cell which, in conjunction with the drying unit, peristaltic pump and autosampler, enables fully automatic operation.

These refractometers are the ideal solution for manufacturers of flavors, fragrances and essential oils, as well as for the pharmaceutical and beverage industries. They are ideal for routine analysis with high sample volumes, quality and purity control of raw materials and end products, or determination of sugar concentration in juices and soft drinks.

Highlights

You are currently viewing a placeholder content from YouTube. To access the actual content, click the button below. Please note that doing so will share data with third-party providers.

More Information

Technical data

Measurement data
DR6000-TF DR6100-TF DR6200-TF DR6300-TF
SCALES Refractive index (nD)
Concentration of sucrose, glucose, fructose and invert sugar (%Brix)
User-defined (unlimited)
MEASUREMENT RANGE nD 1.3200-1.5800
0-95 %Brix
nD 1.3200-1.7000
0-95 %Brix
nD 1.32000-1.58000
0-95 %Brix
nD 1.32000-1.70000
0-95 %Brix
MEASUREMENT ACCURACY 1) nD ±0.0001
±0.1 %Brix
nD ±0.00002
±0.02 %Brix
RESOLUTION nD 0.0001
0.1 %Brix
nD 0.00001
0.01 %Brix
MEASUREMENT PERIOD 3) approx. 4 seconds
MEASUREMENT PRISM Sapphire
LIGHT SOURCE LED 589 nm
SAMPLE VOLUME/FILLING QUANTITY 0.5 ml
COMPONENTS IN CONTACT WITH THE SAMPLE Sapphire, PTFE and stainless steel
Control & device data
OPERATION Stand-alone device (incl. firmware): 7.0 inch capacitive touchscreen, 800 x 480 pixels;
Control also possible with PC keyboard, mouse, barcode scanner (several languages)
DATA INTEGRITY Firmware, data acquisition and storage are compliant with the requirements of 21 CFR Part 11 and EU GMP Annex 11
Conformity with GMP/GLP, pharmacopoeias (USP, BP, JP, Ph. Eur.), FDA, ISO, HACCP, OIML, ASTM, ICUMSA, NIST
ELECTRONIC INTERFACE 1x USB, 1x RS-232, 1x Ethernet, LIMS
Easy integration into networks (DHCP client) or LIMS
ADJUSTMENT Easy-to-understand, menu-guided adjustment.
Certified and traceable calibration standards available as consumables
Temperature Data
TEMPERATURE CONTROL Integrated Peltier sample temperature control system
TEMPERATURE MEASUREMENT RANGE 2) 10-80 °C
TEMPERATURE MEASUREMENT ACCURACY ±0,1 °C
TEMPERATURE COMPENSATION Can be activated (ICUMSA or freely definable)
TEMPERATURE MEASUREMENT With integrated temperature sensor Pt100
TEMPERATURE MEASUREMENT RANGE 10-80 °C
TEMPERATURE MEASUREMENT ACCURACY ±0,1 °C
TEMPERATURE MEASUREMENT RESOLUTION 0,1 °C
Devices Basic Data
HOUSING Aluminum, powder-coated; protection class: IP20
Dimensions (W x H x D) 215 mm x 150 mm x 345 mm
WEIGHT 5 kg
OPERATING VOLTAGE 90-240 V
POWER CONSUMPTION (measurement mode) 40 W
RATED FREQUENCY 47-63 Hz

¹) Under normal conditions for measurement refractive index (λ = 589 nm, 20 °C, 1013 hPa, 50 % relative humidity)

²) For non-condensing ambient conditions

³) After temperature equalization