Flame photometer - Automatic unit with dilution

FP8700

Flame photometer FP8700

Flame photometer FP8700

Calibration standards and cleaning solutions for flame photometers

Calibration standards and cleaning solutions for flame photometers

Flame photometer accessories

Flame photometer accessories

Key features
Applications
Samples
Key features

Automated process for operation with high sample throughput

The flame photometer FP8700 is the laboratory version supplemented by a rotation sampler, diluter, and mixer. Ideal for applications with high sample volumes and high concentrations requiring dilution. Various safety mechanisms ensure particularly high operational safety. Especially control laboratories and testing offices benefit from this independent operation. With the models of the FP8000 series, A.KRÜSS Optronic presents to this day unique measuring devices for the simultaneous measurement of the elements Na, K, Ca and Li. Five measuring channels are available for this, which can be individually configured. Thanks to proven measuring methods and modern device electronics, our flame photometers achieve high-precision and reproducible measurement results and are a simple and particularly cost-effective alternative to analysis techniques such as ICP or AAS.

  • Reliable, simultaneous measurement of up to five alkaline and alkaline earth elements
  • Automatic dilution
  • Automatic sample feed
  • Suitable for high sample throughput and small sample volumes
  • Little care needed by users
  • Very high precision
  • Maximum operational safety through intelligent security mechanisms
  • User administration with two authorization levels
  • Full traceability of the measurement results
  • Extensive interfaces and convenient data transfer
  • Compliance with international standards such as GMP / GLP and 21 CFR Part 11
Applications

Typical applications

  • Quality control in the pharmaceutical, chemical, beverage, food, salt and construction industries
  • Evidence
  • Food laboratories, testing laboratories and testing offices
  • Chemical development

An overview of typical applications with the flame photometer FP8000 series can be found in our campus article on flame photometer applications.

Samples

Samples and substances often analyzed with a flame photometer

  • Measuring potassium and sodium in salt
  • Measurement of sodium, potassium and calcium in various food samples
  • Measurement of sodium, potassium and calcium in fruits and vegetable juices
  • High-purity water
  • Measurement of sodium, potassium and calcium in pharmaceutical reagents and infusion solutions (Ringer’s solution, NaCl solution)
  • Wastewater control by determination of potassium, sodium and magnesium content
  • Concentration determinations of sodium and potassium in fertilizer production
  • Measurement of sodium in glassmaking
  • Review of sodium, potassium or calcium levels in cement production

Model specifications

FP8700NaKLiCa
MEASURING RANGE0.1-45000 ppm
0.004-2000 mmol/l
0.1-45000 ppm
0.003-1100 mmol/l
0.1-45000 ppm
0.014-6000 mmol/l
5.0-45000 ppm
0.125-110 mmol/l
DETECTION LIMIT0.1 ppm
0.0004 mmol/l
0.1 ppm
0.0003 mmol/l
0.1 ppm
0.0014 mmol/l
0.3 ppm
0.0075 mmol/l
PRECISION0.6 % at 40 ppm
0.6 % at 1.74 mmol/l
0.6 % at 40 ppm
0.6 % at 1.03 mmol/l
0.6 % at 40 ppm
0.6 % at 5.71 mmol/l
0.6 % at 40 ppm
0.6 % at 1.00 mmol/l
ACCURACY1 % at 40 ppm
1 % at 1.74 mmol/l
1 % at 40 ppm
1 % at 1.03 mmol/l
1 % at 40 ppm
1 % at 5.71 mmol/l
1 % at 40 ppm
1 % at 1.00 mmol/l
CALIBRATIONlinear with 2 standards
non-linear with 6-8 standards,
cubic approximation
DRIFT1% in 60 min
REFERENCELithium guideline with 35 mg/l or 5 mmol/l
Caesium guideline on request
SAMPLE VOLUME0.25 ml
COMBUSTION GASPropane, Butane or LPG: Recommended for alkaline elements
Acetylene for flame photometry: Recommended for alkaline earth elements
DISPLAYTFT-Display with integrated 8,4“ TFT touchscreen, 800x600 Pixel
INTERFACES2 x USB
1 x Ethernet
1 x RS-232RS-232 for printer
EQUIPMENTBasic unit +
External automation system, consisting of: Sampler, Diluter and Mixer
ACCESSORIESExternal automation system, consisting of: Sampler, Diluter and Mixer
UPGRADEABILITYRetrofit kits to other models available

Technical data flame photometers (PDF, 128 KB)

Requirement profile

To enable us to adapt the flame photometer optimally to your requirements, we need a requirement profile in which you specify your main individual specifications (Samples, concentration, automation, etc.). A binding delivery can only take place after receipt of the fully completed profile.

PDF form requirement profile flame photometer (PDF, 824 KB)

Economical by automatic measurement including dilution

The samples are diluted and measured. The procedure needs virtually no care by the user.

 

Unique safety concept

Automated shutdown of gases

  • At extinguished flame
  • With regular shutdown of the system
  • In case of power failure
  • At low fuel gas, oxidant or cooling air pressure
  • At low fuel gas or air pressure or failure of the cooling air

The safety functions are monitored by an independent safety circuit

 

Intuitive operation

Our flame photometers are easy and intuitive to operate with a large touch screen display and USB mouse.

  • User administration with two authorisation levels
  • Traceability of all measurement data and device-specific data
  • Full traceability of the measurement results
  • Extensive interfaces and convenient data transfer
  • Compliance with international standards such as GMP / GLP and 21 CFR Part 11
 

Calibration, samples, controls and standards

Aqueous samples, controls and standards are measured manually or automatically. The calibration is done via the standards. The measurement results can be checked with the controls.
The emission values and concentration values of the calibration are displayed graphically and in tabular form for the check.

 

Installation

During installation, we include your work environment and processes so that our devices can achieve the best measurement results. Our technical staff ensures that your internal and external quality standards are adhered to. We issue test reports and quality assurance documents and train your employees on our devices.

read more
 

Any Questions?

We will be glad to help.

A.KRÜSS Optronic GmbH
Alsterdorfer Str. 276-278
22297 Hamburg | Germany

Tel. +49 40 5143 17-0
Fax +49 40 5143 17-60

E-Mail sales@kruess.com