Gas analyser for O2 und CO2 Concentration

MAT1500

Gas analyser MAT1500

Gas analyser MAT1500

Gas analyser handle with cannula

Gas analyser handle with cannula

Calibration and test gases for gas analysers

Calibration and test gases for gas analysers

Calibration and test gases for gas analysers with bag and certificate

Calibration and test gases for gas analysers with bag and certificate

Key features
Applications
Key features

The all-rounder

Fitted with a zirconium dioxide sensor and a nondispersive infrared sensor, the MAT1500 is our number one when it comes to the range of application, performance and precision. The measurements are completed within only approx. ten seconds. As is the case with the MAT1400,  highly precise results can be achieved with single measurements of oxygen concentrations thanks to the min/max detection even if only a small sample volume is available.

  • Easy and intuitively to use via touch-screen – even by untrained personnel
  • Low-drift zirconium dioxide sensor (ZrO2in the measuring range from 0 to 100 vol.-%
  • Low-drift non-dispersive infrared sensor (NDIR) measures CO2 in the measuring range from   0 to 100 vol.% with an accuracy of ± 0.5 vol.%
  • Short measurement period
  • Only a small sample volume required
  • Interface package for convenient transfer of measured values (USB, Ethernet, RS-232)
  • Any number of measurement methods for the monitoring of the measurement process according to method, batch, product and/or production line incl. limit value monitoring
Applications

Typical applications

Because of its excellent accuracy of ±0.001 volume per cent with the lowest O2 concentrations, the MAT1400 with its zirconium dioxide sensor is primarily used to test oxygen free modified atmospheres for sterilised finished products or electronic, pharmaceutical or cosmetic products. Thanks to its wide measurement range between 0 and 100 volume per cent, it can also be used for high oxygen concentrations as required for red fresh meat. The non-dispersive infrared sensor with its wide measurement range between 0 and 50 volume per cent  is suitable for most applications of carbon dioxide in modified atmospheres. Its high accuracy of ±0.5 volume per cent CO2 always ensures precise measurement results. A typical application of the MAT1500 is the control of oxygen and carbon dioxide concentration in the following products:

  • Fresh meat
  • Sausages
  • Raw fish
  • Shellfish and molluscs
  • Poultry
  • Carbonated soft drinks
  • Convenient

A comprehensive overview of typical applications of the gas analyzers of the MAT1000 series can be found in our campus article on gas analyzer applications.

Model specifications

 MAT1500
MAIN SPECIFICATIONS
SCALESOxygen
Carbon dioxide
Nitrogen
Any number of user-defined scales
OXYGEN (O2)
MEASUREMENT RANGE0–100 vol%
ACCURACY±0.001 vol% (< 1 vol%)
±0.01 vol% (< 6 vol%)
±0.1 vol% (< 35 vol%)
±0.4 vol% (> 35 vol%)
RESOLUTION0.001 vol%
CARBON DIOXIDE(CO2)
MEASUREMENT RANGE0–100 vol%
ACCURACY±0.5 vol%
RESOLUTION0.1 Vol.-%
NITROGEN (N2)
ARITHMETIC DETERMINATIONResidual content of the gas mixture
O2-SENSOR
TYPEZirconium dioxide sensor ((ZrO2)
MEASUREMENT PRINCIPLEPotentiometric measurement
SERVICE LIFEUnlimited
CROSS-SENSITIVITYCross-sensitive toward inflammable gases
DRIFTDrift-free
CO2-SENSOR
TYPETwo-channel NDIR sensor
MEASUREMENT PRINCIPLEInfrared spectroscopy
SERVICE LIFEUnlimited
CROSS-SENSITIVITYNot affected by moisture
DRIFTLow drift < 1 %/month

Technical data Gas analysers (PDF, 218 KB)

A strong performance package

The gas analysers are robust all-rounders. As a user, you enjoy many benefits such as simple and intuitive operation, intelligent user management, complete data logging, and compliance with global standards.

 

High operating comfort

  • State-of-the-art touch-screen display
  • Uniform operation of all A.KRÜSS laboratory devices
  • Easy, menu-guided adjustment
  • Can be connected to a barcode scanner
  • Printer configuration with numerous options
  • selection of six languages (de, en, fr, es, it, pt)
 

Create your own methods

  • Create any number of methods so that the respective desired parameters can be used for each sample
  • Available method parameters: scales, measurement procedures, limit values, comment and much more
  • Selectable measurement modes: single or continuous
  • Several single measurement procedures: manual, automatic, min/max detection
 

The main screen: Everything relevant at a glance

  • Freely assignable shortcut keys for the most important functions
  • Measured values can be viewed at a glance
  • Optional secondary and tertiary scale to display up to three measured values
  • Method selection with only two clicks
 

Intelligent user administration

  • Two authorisation levels
  • Can be activated or deactivated depending on the requirements
  • Optional setup of user profiles
  • Differentiations with the help of user-defined colour schemes
  • Customised settings for different users or work groups
 

Complete documentation of measured values

  • Integrated measured data storage: complete recording of all measured data as well as system or method settings
  • Consecutive numbering of the measurements
  • Documentation of all measured values of the last 999 measurements
 

Customise your result reports

  • Customisable result table using the column configuration
  • Selection of the results with filters
  • Ascending or descending sorting of the datasets according to number
 

Integrated interfaces, flexible data export

  • Print-out on serial ASCII printer at RS-232 interface
  • Print-out on network printer in PDF or GS format
  • Print-out as PDF on USB flash drive or to network share
  • Export of the measurement results in HTML or CSV format on USB flash drive or to network share
  • Easy integration into existing networks (DHCP client) or a LIMS
 

The right sensors for gas analysis

Explanation of the electrochemical oxygen cell (EC) and the zirconium dioxide sensor (ZrO2) for the O2-measurement as well as the non-invasive infrared sensor (NDIR) for the CO2-measurement.

Sensor overview
 

Any Questions?

We will be glad to help.

A.KRÜSS Optronic GmbH
Alsterdorfer Str. 276-278
22297 Hamburg | Germany

Tel. +49 40 5143 17-0
Fax +49 40 5143 17-60

E-Mail sales@kruess.com