Fully automatic measurements

The M5000 provides fast, easy and automatic examination of powdery substances with a melting point up to 400 °C. Due to the automatic determination of the melting point, an objective measurement result is always guaranteed. The high preheating rate and the integrated fan cooling allow fast and reliable measurements in any temperature range. The display provides a well-arranged readout of all important measuring data.

Your advantages

  • Fully automatic measurement
  • 1 capillary intake
  • LCD display
  • Alarm signal when reaching the melting point
  • Fast cooling through integrated fan
  • Digital display of all important data
  • Display in German or English
  • With RS-232 interface for printer CBM910
  • Easy-to-clean membrane keypad
  • Includes protective cover and 100 capillaries
  • Small sample volume
Applications

Typical applications

The determination of the melting point is of great importance in the identity test, since many substances can be determined by their melting point.

The purity of substances can also be qualitatively measured via melting point. A well-defined melting point is only given for 100 percent pure substances. Due to impurities, such as impurities or dopants, most substances have a melting interval (beginning of the melt to completion of the melt) of a few degrees Celsius. Even impurities during the manufacture of chemicals cause this melting point to expand to a melting interval. As a result, most substances that are examined with a melting point meter have a melting interval. This effect is used to identify the degree of impurity of the substance.

Model specifications

 M5000
MEASUREMENT RANGE25–400 °C
MEASUREMENT ACCURACY±0.3 °C (25–200 °C)
±0.5 °C (200–400 °C)
RESOLUTION0.1 °C
PREHEATING RATETo 200 °C approx. 4.0 min
To 300 °C approx. 9.0 min
HEATING RATE1 °C min-1
NO. OF CAPILLARIES1
CAPILLARY Ø1.4 mm
INTERFACERS-232
IP CODEIP20
POWER SUPPLY90–264 V
DIMENSIONS (W X H X D)220 mm x 150 mm x 340 mm
WEIGHT4.1 kg

Why should I invest in an automated melting point meter?

melting point

The modern melting point meter is fast and accurate, providing reproducible results which do not rely on the skill of the individual operator. Measurements can be printed for a permanent record. Digital control means that the sample is heated, the melting point recorded, and the apparatus cooled ready for the next sample is just a few minutes. Sample sizes are small, being just enough powdered material to fill a small capillary tube, and the output reading is accurate to within 0.3 – 0.5C with good reproducibility.

More in the application video

Service

Application Consulting

We will be happy to assist you with product and application questions - by telephone, on site or in our training center in Hamburg. Whether it's advice and support, quick problem solving through application experience, help with system automation and integration, questions about industry standards, our experts are there for you.

read more
 

Any Questions?

We will be glad to help.

A.KRÜSS Optronic GmbH
Alsterdorfer Str. 276-278
22297 Hamburg | Germany

Tel. +49 40 5143 17-0
Fax +49 40 5143 17-60

E-Mail sales@kruess.com